Product Details:
Payment & Shipping Terms:
|
Bluetooth: | 4.0 Version For IOS And Andriod | Color: | Orange, Silver |
---|---|---|---|
Packing: | Retail Giftbox Pack | Display Screen: | OLED Screen Display With Curve |
Spo2 Measure Range: | 35~99% | Use Application: | Home, Clinc, Hospital, Drugstore |
High Light: | Fingertip Pulse Oximeter,portable finger pulse oximeter |
CE Bluetooth 4.0 Finger Pulse Oximeter SPO2 meter with OLED display
1.Specification:
1.1 Measurement Range
Spo2: 0 -100%
Pulse Rate: 25 ~ 250 BPM
1.2 Resolution
Spo2: 1%
Pulse Rate: 1BPM
1.3 Accuracy
SPO2 Accuracy: 2% (70-100%)
No defined (35-70%)
Pulse Rate 1% (25-250BPM)
Input Voltage:3V
Input Current: <40mA
2.Environment
2.1 Temperature
Working:0 ~ 45°C
Storage:-20~55°C
2.2 Humidity
Working:30~95%
Storage:10~100%
3.Standards
EN865 Calibration specification for pulse oximeters
EN60601-1 Safety for medical equipment
EN60601-1-4 General safety standard
Precautions for use:
1.Please use two 5 AAA alkaline batteries or rechargeable batteries, do not use carbon or inferior battery.
2 . Please take out of the battery, if long time no use.
3 . Please timely replace batteries if low batteries indicator. Because battery supply deficiency may influence the measurement accuracy.
4.Please make sure the correct polarity direction to install battery, otherwise unable to boot.
Functions:
Common troubleshooting:
Electric indicator:
Safety classification: Medical electrical equipment protection according to the taxonomy of the genus the instrument for the II type BF equipment.
Power supply for DC 3 V.
Working environment, transportation and storage:
1. Normal working conditions
A ) Enviroment temperature: 5°C ~ 40 °C
B) Relative humidity:≤ 80%.
C ) Atmospheric pressure: 86kPa to 106kPa.
D)Storage
Good packed oximeters should be stored in a temperature of -40 ~ +70 °C, relative humidity 10%~95%, no corrosive gas and good ventilation indoor.
Operation Instructions:
PC Software
1. Turn on
Opening the computer or mobile Bluetooth function, Searching Bluetooth oximetry , and pair, input pair code: 0000. The last step is to connected by Bluetooth serial port.
2. Opening the random carring software spo2.exe
3. According to a Bluetooth serial number to choose a serial port, and click the start button to start the Oxygen detection function.
Android App
1. Open “Spo2 Mes” app.
2. Open Bluetooth, search the device, then enter the pairing code: “0000” or “1234”.